Outcomes of Directional Branches of the T-Branch Off-the-Shelf Multi-Branched Stent-Graft

Link:
Autor/in:
Erscheinungsjahr:
2022
Medientyp:
Text
Beschreibung:
  • BACKGROUND: A controversy on bridging covered stent (BCS) choice, between self-expanding (SECS) and balloon-expandable (BECS) stents, still exists in branched endovascular repair. This study aimed to determine the primary target vessel (TV) patency in patients treated with the t-Branch device and identify factors impairing the outcomes.

    METHODS: A retrospective study was undertaken, including patients treated with the t-Branch (Cook Medical, Bloomington, IN, USA) between 2014 and 2019 (early 2014-2016; late 2017-2019). The endpoint was the primary patency (CT: celiac trunk, SMA, superior mesenteric artery, RRA: right renal artery, LRA: left renal artery) during the follow-up. Any branch instability event was assessed. The factors affecting the patency were determined using multivariable regression models and Kaplan-Meier analyses.

    RESULTS: In total, 2018 TVs were analyzed; 1542 SECSs and 476 BECSs. The CT patency was 99.8% (SE 0.2%) at the 1st month, with no other event. The SMA patency was 97.8% (SE 1) at the 12th month. The RRA patency was 96.7% (SE 2) at the 24th month. The LRA patency was 99% (SE 0.4) at the 6th month. Relining was the only factor independently associated with the SMA patency (OR 8.27; 95% CI 1.4-4.9; p = 0.02). The freedom from instability was 62% (SE 4.3%) and 45% (SE 5.4%) at the 24th month and 36th month. No significant difference was identified between the BECSs and SECSs in the early or late experience.

    CONCLUSION: BCS for the t-Branch branches performed with a good primary patency during the short-term follow-up. The type of BCS did not influence the patency. Relining might be protective for SMA patency.

Lizenz:
  • info:eu-repo/semantics/openAccess
Quellsystem:
Forschungsinformationssystem des UKE

Interne Metadaten
Quelldatensatz
oai:pure.atira.dk:publications/afa598dd-b26e-42b1-9abb-788bfe1f5724