MRI-based quantitative susceptibility mapping (QSM) and R2* mapping of liver iron overload Comparison with SQUID-based biomagnetic liver susceptometry

Link:
Autor/in:
Erscheinungsjahr:
2017
Medientyp:
Text
Beschreibung:
  • PURPOSE: We aimed to determine the agreement between quantitative susceptibility mapping (QSM)-based biomagnetic liver susceptometry (BLS) and confounder-corrected R2* mapping with superconducting quantum interference device (SQUID)-based biomagnetic liver susceptometry in patients with liver iron overload.

    METHODS: Data were acquired from two healthy controls and 22 patients undergoing MRI and SQUID-BLS as part of routine monitoring for iron overload. Magnetic resonance imaging was performed on a 3T system using a three-dimensional multi-echo gradient-echo acquisition. Both magnetic susceptibility and R2* of the liver were estimated from this acquisition. Linear regression was used to compare estimates of QSM-BLS and R2* to SQUID-BLS.

    RESULTS: Both QSM-BLS and confounder-corrected R2* were sensitive to the presence of iron in the liver. Linear regression between QSM-BLS and SQUID-BLS demonstrated the following relationship: QSM-BLS = (-0.22 ± 0.11) + (0.49 ± 0.05) · SQUID-BLS with r(2)  = 0.88. The coefficient of determination between liver R2* and SQUID-BLS was also r(2)  = 0.88.

    CONCLUSION: We determined a strong correlation between both QSM-BLS and confounder-corrected R2* to SQUID-BLS. This study demonstrates the feasibility of QSM-BLS and confounder-corrected R2* for assessing liver iron overload, particularly when SQUID systems are not accessible. Magn Reson Med, 2016. © 2016 International Society for Magnetic Resonance in Medicine.

Lizenz:
  • info:eu-repo/semantics/restrictedAccess
Quellsystem:
Forschungsinformationssystem des UKE

Interne Metadaten
Quelldatensatz
oai:pure.atira.dk:publications/dc92b91b-3d4d-428c-a34c-881c53b8f65b