Optimisation of inductive telemetry links in an implantable medical device for a wireless range of 25 cm

Link:
Autor/in:
Verlag/Körperschaft:
Hamburg University of Technology
Erscheinungsjahr:
2018
Medientyp:
Text
Schlagworte:
  • 000: Allgemeines, Wissenschaft
  • 000
Beschreibung:
  • The optimisation of inductive telemetry link for increasing wireless range of a medical implant device is studied in detail. An active implant device using separate wireless power and data links at 2 MHz and 420 kHz has been fabricated and tested. A test setup consisting of RF power transmitter and data reader hardware has been implemented. Measurement results show that the fabricated device achieved a wireless range of 25 cm.
Beziehungen:
DOI 10.1049/el.2018.6074
Quellsystem:
TUHH Open Research

Interne Metadaten
Quelldatensatz
oai:tore.tuhh.de:11420/2695