Direct percutaneous transaxillary implantation of a novel self-expandable transcatheter heart valve for aortic stenosis

Link:
Autor/in:
Erscheinungsjahr:
2017
Medientyp:
Text
Beschreibung:
  • OBJECTIVES: The aim of this study was to evaluate safety, feasibility, and efficacy of transaxillary TAVI using a novel self-expandable transcatheter heart valve (THV) via a direct percutaneous technique.

    METHODS: The analysis comprises 12 consecutive patients with severe aortic stenosis treated with the ACURATE Neo™ THV (Symetis) by direct percutaneous transaxillary access.

    RESULTS: Patients presented with a mean age of 79.9 ± 6.2 years and high calculated surgical risk (logEuroSCORE I: 19.55 ± 15.00%; STS PROM score: 5.42 ± 3.25%). Transaxillary TAVI was performed in 11/12 via the left and in 1/12 via the right-sided axillary artery. Device success was achieved in all patients (12/12). Mortality rate at 30 days was 0% (0/12). No major and one minor access site complication (1/12) occurred. No major or life-threatening access site-related bleeding resulted. None of the patients suffered from postprocedural stroke/transient ischemic attack. There was no need for a permanent pacemaker. Mean effective orifice area post implantation was 2.1 ± 0.29 cm2 and mean aortic gradient 6.2 ± 1.5 mm Hg. In none of the patients, greater than moderate aortic regurgitation or paravalvular leakage was detected at discharge (0/12).

    CONCLUSIONS: Our study suggests that the implantation of the second-generation self-expandable ACURATE Neo™ via a transaxillary access is technically feasible and safe and yields excellent short-term clinical results. © 2017 Wiley Periodicals, Inc.

Lizenz:
  • info:eu-repo/semantics/restrictedAccess
Quellsystem:
Forschungsinformationssystem des UKE

Interne Metadaten
Quelldatensatz
oai:pure.atira.dk:publications/6fff2dd4-296a-46e9-8a29-2209d3c83bfa