Valve-in-valve-in-valve:Balloon expandable transcatheter heart valve in failing self-expandable transcatheter heart valve in deteriorated surgical bioprosthesis

Link:
Autor/in:
Erscheinungsjahr:
2018
Medientyp:
Text
Beschreibung:
  • Valve-in-valve (ViV) procedures for failing bioprostheses carry a certain risk for device malfunction. We herein report a case of a failing Evolut R in a deteriorated Mitroflow, treated with a Sapien 3. An 81 year old female patient received surgical aortic valve replacement and was treated by ViV due to deterioration. Three years later, echocardiography revealed a pressure gradient of peak/mean 105/63 mmHg. Subsequently, a second ViV procedure with initial intentional rupture of the bioprosthetic stent was performed. Immediate stent recoil of the Evolut R prompted implantation of a Sapien 3. In 30-day follow-up, mean pressure gradient of 30 mmHg and nearly complete symptom relief was documented. Fracture of a surgical bioprosthetic stent is feasible in a ViV configuration. Supra-annular placement of a balloon-expandable THV as ViV-in-valve is feasible with suboptimal hemodynamic results in this case. Risk of re-do surgery should be weighted against anticipated hemodynamic and clinical results.

Lizenz:
  • info:eu-repo/semantics/restrictedAccess
Quellsystem:
Forschungsinformationssystem des UKE

Interne Metadaten
Quelldatensatz
oai:pure.atira.dk:publications/b2d157fe-0277-457d-9d19-f64a119adf9b