Wireless blood pressure measurement implant electronics for integration in a stent craft

Link:
Autor/in:
Verlag/Körperschaft:
Hamburg University of Technology
Erscheinungsjahr:
2016
Medientyp:
Text
Schlagworte:
  • telemetry
  • wireless power transmission
  • inductive coupling
  • Implantat
  • pressure measurement
  • 620: Ingenieurwissenschaften
  • 620
Beschreibung:
  • This paper describes the design, implementation and testing of a wireless blood pressure sensing implant. The implant is mounted on a stent graft. The implant along with the stent graft must fit into a 16F applicator stent delivery system. The implant is powered using the energy harvested from wireless power transmission. The measured pressure data is sent wirelessly to an external data reader hardware. The pressure measurement must have resolution of 1 mmHg.