Combined preclinical magnetic particle imaging and magnetic resonance imaging: Initial results in mice

Link:
Autor/in:
Verlag/Körperschaft:
Hamburg University of Technology
Erscheinungsjahr:
2015
Medientyp:
Text
Schlagworte:
  • co-registration
  • magnetic particle imaging
  • MR imaging
  • preclinical
  • superparamagnetic iron oxide
  • workflow
Beschreibung:
  • Purpose: Magnetic particle imaging (MPI) is a new radiologic imaging modality. For the first time, a commercial preclinical scanner is installed. The goal of this study was to establish a workflow between MPI and magnetic resonance imaging (MRI) scanners for a complete in vivo examination of a mouse and to generate the first co-registered in vivo MR-MP images. Materials and Methods: The in vivo examination of five mice were performed on a preclinical MPI scanner and a 7 Tesla preclinical MRI system. MRI measurements were used for anatomical referencing and validation of the injection of superparamagnetic iron oxide (SPIO) particles during a dynamic MPI scan. We extracted MPI data of the injection phase and co-registered it with MRI data. Results: A workflow process for a combined in vivo MRI and MPI examination was established. A successful injection of ferucarbotran was proven in MPI and MRI. MR-MPI co-registration allocated the SPIOs in the inferior vena cava and the heart during and shortly after the injection. Conclusion: The acquisition of preclinical MPI and MRI data is feasible and allows the combined analysis of MR-MPI information.